In-Situ Systematic Error Correction for Digital Volume Correlation Using a Reference Sample

B. Wang, B. Pan*, G. Lubineau

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

7 Scopus citations

Abstract

The self-heating effect of a laboratory X-ray computed tomography (CT) scanner causes slight change in its imaging geometry, which induces translation and dilatation (i.e., artificial displacement and strain) in reconstructed volume images recorded at different times. To realize high-accuracy internal full-field deformation measurements using digital volume correlation (DVC), these artificial displacements and strains associated with unstable CT imaging must be eliminated. In this work, an effective and easily implemented reference sample compensation (RSC) method is proposed for in-situ systematic error correction in DVC. The proposed method utilizes a stationary reference sample, which is placed beside the test sample to record the artificial displacement fields caused by the self-heating effect of CT scanners. The detected displacement fields are then fitted by a parametric polynomial model, which is used to remove the unwanted artificial deformations in the test sample. Rescan tests of a stationary sample and real uniaxial compression tests performed on copper foam specimens demonstrate the accuracy, efficacy, and practicality of the presented RSC method.

Original languageEnglish (US)
Pages (from-to)427-436
Number of pages10
JournalExperimental Mechanics
Volume58
Issue number3
DOIs
StatePublished - Mar 1 2018

Keywords

  • Digital volume correlation
  • Reference sample compensation
  • Self-heating effect
  • Systematic error

ASJC Scopus subject areas

  • Aerospace Engineering
  • Mechanics of Materials
  • Mechanical Engineering

Fingerprint

Dive into the research topics of 'In-Situ Systematic Error Correction for Digital Volume Correlation Using a Reference Sample'. Together they form a unique fingerprint.

Cite this