Abstract
Data independent acquisition–mass spectrometry (DIA–MS) is becoming widely utilised for robust and accurate quantification of samples in quantitative proteomics. Here, we describe the systematic evaluation of the effects of DIA precursor mass range on total protein identification and quantification. We show that a narrow mass range of precursors (~250 m/z) for DIA–MS enables a higher number of protein identifications. Subsequent application of DIA with narrow precursor range (from 400 to 650 m/z) on an Arabidopsis sample with spike-in known proteins identified 34.7% more proteins than in conventional DIA (cDIA) with a wide precursor range of 400–1200 m/z. When combining several DIA–MS analyses with narrow precursor ranges (i.e., 400–650, 650–900 and 900–1200 m/z), we were able to quantify 10,099 protein groups with a median coefficient of variation of
Original language | English (US) |
---|---|
Pages (from-to) | 982 |
Journal | Life (Basel, Switzerland) |
Volume | 11 |
Issue number | 9 |
DOIs | |
State | Published - Sep 28 2021 |
ASJC Scopus subject areas
- Space and Planetary Science
- Palaeontology
- General Biochemistry, Genetics and Molecular Biology
- Ecology, Evolution, Behavior and Systematics