T1 - Performance characteristics of six intact parathyroid hormone assays
AU - La'ulu, Sonia L.
AU - Roberts, William L.
N1 - Generated from Scopus record by KAUST IRTS on 2023-09-20
PY - 2010/12/1
Y1 - 2010/12/1
N2 - The aim of this study was to evaluate the performance characteristics of 6 intact parathyroid hormone assays: Access 2 (Beckman Coulter, Fullerton, CA), ARCHITECT i2000SR (Abbott Diagnostics, Abbott Park, IL), ADVIA Centaur (Siemens Healthcare Diagnostics, Deerfield, IL), Modular E170 (Roche Diagnostics, Indianapolis, IN), IMMULITE 2000 (Siemens Healthcare Diagnostics), and LIAISON (DiaSorin, Stillwater, MN). Sample collection tubes and storage conditions were compared. Imprecision studies were performed using commercial quality control materials. Linearity was assessed using pools prepared from samples. For method comparison, serum and EDTA plasma samples were tested by all methods, and the ARCHITECT was used as the comparison method. Reference intervals were determined using various vitamin D cutoffs. The types of collection tubes and storage conditions are more important for some methods than others. Total coefficients of variation were 10.9% or less. The maximum deviation from the target recovery for linearity ranged from 5.0% to 82.2%. Bland-Altman plots demonstrated percentage biases ranging from -36.3% to 24.4%. The lower limit of the reference interval was not influenced by vitamin D status, whereas the upper reference limit was affected. © American Society for Clinical Pathology.
AB - The aim of this study was to evaluate the performance characteristics of 6 intact parathyroid hormone assays: Access 2 (Beckman Coulter, Fullerton, CA), ARCHITECT i2000SR (Abbott Diagnostics, Abbott Park, IL), ADVIA Centaur (Siemens Healthcare Diagnostics, Deerfield, IL), Modular E170 (Roche Diagnostics, Indianapolis, IN), IMMULITE 2000 (Siemens Healthcare Diagnostics), and LIAISON (DiaSorin, Stillwater, MN). Sample collection tubes and storage conditions were compared. Imprecision studies were performed using commercial quality control materials. Linearity was assessed using pools prepared from samples. For method comparison, serum and EDTA plasma samples were tested by all methods, and the ARCHITECT was used as the comparison method. Reference intervals were determined using various vitamin D cutoffs. The types of collection tubes and storage conditions are more important for some methods than others. Total coefficients of variation were 10.9% or less. The maximum deviation from the target recovery for linearity ranged from 5.0% to 82.2%. Bland-Altman plots demonstrated percentage biases ranging from -36.3% to 24.4%. The lower limit of the reference interval was not influenced by vitamin D status, whereas the upper reference limit was affected. © American Society for Clinical Pathology.
UR - https://academic.oup.com/ajcp/article-lookup/doi/10.1309/AJCPLGCZR7IPVHA7
UR - http://www.scopus.com/inward/record.url?scp=78649678926&partnerID=8YFLogxK
U2 - 10.1309/AJCPLGCZR7IPVHA7
DO - 10.1309/AJCPLGCZR7IPVHA7
M3 - Article
SN - 0002-9173
VL - 134
SP - 930
EP - 938
JO - American Journal of Clinical Pathology
JF - American Journal of Clinical Pathology
IS - 6
ER -